Design of micro-sampling system with high precision for small enzyme immunoassay analyzer |
SHANG Zhi-wu, ZHOU Xiang-ping, LI Cheng |
Tianjin Modern Electromechanical Equipment Technology Key Laboratory, Tianjin Polytechnic University, Tianjin 300387, China |
|
|
Abstract To improve the reliability and accuracy of the automatic enzyme immunoassay analyzer's micro-sampling system, a small autoinjection system was developed. A sophisticated injection arm was designed by the means of dislocation parallel distribution of the screw and injector piston rod. It possessed the function of pipetting, taking and removing the pipette tips. In the control system STM32 controller was used, controlling the single-axis S-type acceleration/deceleration algorithm and multithreaded coordinated motion. The acceleration/deceleration curves were analyzed and optimized by using the method of segmentation, the minimum injection rate of 1 μL was realized and the step rate was 0.05 μL. The position accuracy and the sampling precision of the injection system were verified by experiments, and the least squares linear fitting method was used to compensate the system sampling error. The results showed that the optimized S-type acceleration and deceleration algorithm could improve the motion characteristics of the stepping motor, avoiding the out-of-step and overshoot effectively, and the injection arm could reach a reliable position accuracy. The microinjection system had higher accuracy and stability after error compensation, at the test point 10, 50, 100 μL, the injection precision increased respectively from ±7.2%, ±5.3%, ±3.2% to ±1.8%, ±1.28%, ±1.15%. The system meets the miniaturization and high-precision design requirements, which has good application and popularization value.
|
Received: 03 July 2017
Published: 28 October 2018
|
|
高精度小型酶联免疫分析仪微量进样系统设计
为提高酶联免疫分析仪微量进样的可靠性与精度,自主研发了一种小型微量自动进样系统。采用丝杆与进样器活塞杆错位平行分布的方式,设计了精巧的进样机构,实现移液和取、退吸头的功能;利用STM32核心控制器,实现了单轴的S型加减速控制以及多轴协调、多线程运动控制;利用分段的方法对加减速曲线进行分析与优化,实现系统最小进样量为1 μL,以0.05 μL的进样分辨率步进;通过试验校验了进样臂的位置精度与进样精度,采用最小二乘线性拟合方法对系统进样误差进行补偿校正。研究结果表明:优化后的S型加减速算法改善了步进电机的运动特性,有效避免了失步与过冲现象,使进样机构具有较高的位置精度;误差补偿后的微量进样系统拥有更高的进样精度与稳定性,在检定进样量为10,50,100 μL时的进样精度分别由±7.2%,±5.3%,±3.2%提高到±1.8%,±1.28%,±1.15%,满足仪器小型化、高精度的设计要求,具有良好的实际应用与推广价值。
关键词:
酶联免疫分析仪,
微量进样器,
STM32,
运动控制,
误差补偿
|
|
[[1]] |
TAO L, LONG W, WANG Y Q. Design and implementation of the control system for an ARM9-based biochemical analyzer[J]. Advance Materials Research, 2011, 268/270:454-457.
|
|
|
[[2]] |
刘雄.800速全自动生化分析仪运动控制系统关键技术研究[D].绵阳:西南科技大学计算机科学与技术学院,2015:1-3. LIU Xiong. Research on key technologies about motion control system of 800-speed automatic biochemical analyzer[D]. Mianyang:Southwest University of Science and Technology, School of Computer Technology and Science, 2015:1-3.
|
|
|
[[3]] |
武通园.高速全自动化生化分析仪高精度加样技术研究[D].深圳:深圳大学医学部,2016:1-4. WU Tong-yuan. Research on high accuracy accumulation technology of high speed fully automatic biochemical analyzer[D]. Shenzhen:Shenzhen University, School of Health Science Center, 2016:1-4.
|
|
|
[[4]] |
祝连庆,张文昌,董明利,等.一种提高全自动酶免分析仪微量移液精度的方法[J].仪器仪表学报,2013,34(5):1008-1014. ZHU Lian-qing, ZHANG Wen-chang, DONG Ming-li, et al. Method to improve the micropipette accuracy of full-automatic ELISA system[J]. Chinese Journal of Scientific Instrument, 2013, 34(5):1008-1014.
|
|
|
[[5]] |
CHANG Zhu, SHANG Zhi-wu, LIU Zhen-wu, et al. A small automated enzyme immunoassay analyzer system design and implementation based on immunoenzymatic techniques[J]. Key Engineering Materials, 2016,693:1308-1313.
|
|
|
[[6]] |
常海涛,祝连庆,娄小平,等.一种全自动酶免分析仪移液过程评估新方法[J].仪器仪表学报,2014,35(7):1622-1629. CHANG Hai-tao, ZHU Liang-qing, LOU Xiao-ping, et al. A fully automatic enzyme immunoassay analyzer shift liquid process to evaluate the new method instrument[J]. Chinese Journal of Instrumentation, 2014, 35(7):1622-1629.
|
|
|
[[7]] |
刘镇武,尚志武,黄炎彬.全自动生化进样系统的设计[J].工程设计学报,2016,23(6):612-619. LIU Zhen-wu, SHANG Zhi-wu, HUANG Yang-bin. Design of full automatic biochemical sampling system[J]. Chinese Journal of Engineering Design, 2016, 23(6):612-619.
|
|
|
[[8]] |
林丽文,吕慧.关于全自动酶联免疫分析仪的携带污染分析[J].临床医学研究与实践,2016,1(13):81. LIN Li-wen, LÜ Hui. Analysis of carry contamination in enzyme immunoassay analyzer[J]. Clinical Research and Practice, 2016, 1(13):81.
|
|
|
[[9]] |
杜新杰,祝连庆,娄小平,等.基于模糊PID的酶免分析仪加样臂运动控制[J].电子测量与仪器学报,2014,28(3):327-333. DU Xin-jie, ZHU Lian-qing, LUO Xiao-ping, et al. Fuzzy PID for adding sample arm movement control of ELISA processor[J]. Journal of Electronic Measurement and Instrument, 2014, 28(3):327-333.
|
|
|
[[10]] |
黄兆斌,黄云龙,余世明.几种步进电机加减速方法的对比研究及其应用[J].机电工程,2011,28(8):951-953,974. HUANG Zhao-bin, HUANG Yun-long, YU Shi-ming. Several step stepper motor acceleration and deceleration method of comparative study and its application[J]. Mechanical and Electrical Engineering, 2011, 28(8):951-953, 974.
|
|
|
[[11]] |
杨超,张冬泉.基于S曲线的步进电机加减速的控制[J].机电工程,2011,28(7):813-817. YANG Chao, ZHANG Dong-quan. Control of acceleration and deceleration of stepping motor based on S curve[J]. Mechanical and Electrical Engineering, 2011, 28(7):813-817.
|
|
|
[[12]] |
刘慧英,范宝山.基于STM32的多步进电机控制系统研究[J].测控技术,2010,29(6):54-57. LIU Hui-ying, FAN Bao-shan. Study of control system of multiple step motors based on STM32[J]. Measurement & Control Technology, 2010, 29(6):54-57.
|
|
|
[[13]] |
张团善,肖磊,张娜.基于STM32和FPGA的多轴运动控制器的设计[J].电子测量技术,2011,34(5):61-63,67. ZHANG Tuan-shan, XIAO Lei, ZHANG Na. Design of multiaxial motion controller based on STM32 and FPGA[J]. Electronic Measurement Technology, 2011, 34(5):61-63, 67.
|
|
|
[[14]] |
程鑫.多轴超精密同步运动控制系统中的数据交互机制[J].机械工程学报,2014,50(17):149-156. CHENG Xin. Data interchange mechanism in multi-axis ultra-precise synchronous motion control system[J]. Journal of Mechanical Engineering, 2014, 50(17):149-156.
|
|
|
[[15]] |
龚燕妮.全自动生化分析仪多任务优化调度与检测方法研究[D].长沙:湖南大学自动化学院,2012:31-35. GONG Yan-ni. Research on multitask scheduling optimization and inspection of automatic biochemical analyzer[D]. Changsha:Hunan University, College of Electrical and Information Engineering, 2012:31-35.
|
|
|
[[16]] |
ISO Technical Committees. Piston-operated volumetric apparatus Part6:Gravimetric methods for the determination of measurement error:ISO 8655-6:2002[S]. German:German Institute for Standardization, 2002:12-13.
|
|
|
[[17]] |
刘志刚,万良晨.自动进样器的误差分析[J].广东技术师范学院学报(自然科学),2014(3):38-42. LIU Zhi-gang, WAN Liang-chen. The error analysis of automatic sample injector[J]. Journal of Guangdong Polytechnic Normal University, 2014(3):38-42.
|
|
|
|