Gradient spoiling optimization for phase contrast magnetic resonance sequence" /> 相位对比磁共振序列中的扰相梯度优化
Please wait a minute...
浙江大学学报(工学版)
计算机技术﹑电信技术     
相位对比磁共振序列中的扰相梯度优化
邹云鹏, 康雁
东北大学 中荷生物医学与信息工程学院,辽宁 沈阳 110819
Gradient spoiling optimization for phase contrast magnetic resonance sequence
ZOU Yun-peng, KANG Yan
Sino-Dutch Biomedical and Information Engineering School, Northeastern University, Shenyang 110819, China
 全文: PDF(1130 KB)   HTML
摘要:

针对扰相梯度在基于梯度回波(GRE)的二维相位对比磁共振成像(2D PC-MRI)序列中的作用进行理论分析和实验研究.使用基于流动质子的2D PC-MRI仿真对扰相梯度的作用进行分析;在流动模体和志愿者实验中使用的扰相梯度分别为0π、2π、4π、6π、8π和10π,对成像结果进行测量,得到对应的流量.实验结果表明:当2D PC-MRI序列中的的扰相梯度为0π和2π时,流动模体测量结果与设定值之间存在较大的误差,志愿者实验中扰相梯度为0π和2π的实验组与其他实验组的测量结果相比误差明显;当2DPC-MRI序列中的扰相梯度大于等于4π时,各实验组之间的测量误差小于5%.从保证成像质量和减少成像时间两方面考虑,相位对比磁共振序列中最优的扰相梯度为4π.

Abstract:

Effect of gradient spoiling in gradient recalled echo (GRE) based two-dimensional phase contrast magnetic resonance imaging (2D PC-MRI) sequence was studied by theoretical analysis and experimental research. 2D PC-MRI simulation was applied to analyze the effect of gradient spoiling on flowing protons. Gradient spoilings of 0π,2π,4π,6π,8π and 10π were employed for flow phantom and volunteer experiment, and the relative flow rates through calculation of imaging results were obtained. Results showed that there was obvious error between the measured flow and setting values when the gradient spoilings of 0π and 2π were applied in the flow phantom experiment. There were significant errors comparing the group with gradient spoilings of 0π and 2π to other groups’ results in the volunteer experiment. When the gradient spoiling was equal or greater than 4π, the measuring errors between the experiment groups were less than 5%. However, due to the fact that the greater use of gradient spoiling requires more time for imaging, 4π was the optimal gradient spoiling in 2D PC-MRI sequence when considering the image quality and the imaging time.

出版日期: 2015-06-01
:  R 811  
基金资助:

国家自然科学基金资助项目(61071213,61372014)

通讯作者: 康雁,男,教授     E-mail: kangyan@bmie.neu.edu.cn
作者简介: 邹云鹏(1983—),男,博士生,从事磁共振成像研究. E-mail: y_p_zou@tom.com
服务  
把本文推荐给朋友
加入引用管理器
E-mail Alert
RSS
作者相关文章  

引用本文:

邹云鹏, 康雁. 相位对比磁共振序列中的扰相梯度优化[J]. 浙江大学学报(工学版), 10.3785/j.issn.1008-973X.2015.06.008.

ZOU Yun-peng, KANG Yan.

Gradient spoiling optimization for phase contrast magnetic resonance sequence
. JOURNAL OF ZHEJIANG UNIVERSITY (ENGINEERING SCIENCE), 10.3785/j.issn.1008-973X.2015.06.008.

链接本文:

http://www.zjujournals.com/eng/CN/10.3785/j.issn.1008-973X.2015.06.008        http://www.zjujournals.com/eng/CN/Y2015/V49/I6/1055

[1] BRYANT D J,PAYNE J A,FIRMIN D N,et al. Measurement of flow with NMR imaging using a gradient pulse and phase difference technique [J]. Journal of Computer Assisted Tomography,1984,8(4): 588-593.
[2] THOMPSON R B, MCVEIGH E R. Fast measurement of intracardiac pressure differences with 2D breath-hold phase-contrast MRI[J]. Magnetic Resonance in Medicine,2003,49(6): 1056-1066.
[3] SRICHAI M B, LIM RP, WONG S, et al. Cardiovascular application of phase-contrast MRI[J]. American Journal of Roentgenology,2009,192(3): 662-675.
[4] LOTZ J, MEIER C, LEPPERT A, et al. Cardiovascular flow measurement with phase-contrast MR imaging: basic facts and implementation [J]. Radiographics,2002,22(3): 651-671.
[5] GATEHOUSE P D,KEEGAN J,CROWE L A,et al. Applications of phase-contrast flow and velocity imaging in cardiovascular MRI [J]. European Radiology,2005,15(10): 2172-2184.
[6] BATTAL B,KOCAOGLU M,BULAKBASI N,et al. Cerebrospinal fluid flow imaging by using phase-contrast MR technique[J]. British Journal of Radiology,2011,84(1004):758-765.
[7] OVERALL W R, NISHIMURA D G, HU B S. Fast phase-contrast velocity measurement in the steady state [J]. Magnetic Resonance Imaging,2002,48(5): 890-898.
[8] MOSER K W,GEORGIADIS J G,BUCKIUS R O. On the accuracy of EPI-based phase contrast velocimetry [J]. Magnetic Resonance Imaging,2000,18(9): 1115-1123.
[9] GATEHOUSE P D,KEEGAN J,CROWE L A,et al. Applications of phase-contrast flow and velocity imaging in cardiovascular MRI [J]. European Radiology, 2005, 15(10): 2172-2184.
[10] BERNSTEIN M A,KING K F,ZHOU X J. Handbook of MRI pulse sequences [M]. Burlington: Elsevier, 2004: 349-357.
[11] LEUPOLD J,HENNIG J,SCHEFFLER K. Moment and direction of the spoiler gradient for effective artifact suppression in RF-spoiled gradient echo imaging [J]. Magnetic Resonance Imaging,2008,60(1): 119-127.
[12] LEUPOLD J, HENNIG J. Increasing spoiling efficiency in RF-spoiled gradient echo sequence by averaging of RF phase-cycle-adapted K-spaces [J]. Magnetic Resonance Imaging,2011,66(4): 1123-1128.
[13] MANFRED T, RAIMUND E. Cardiac imaging: a multimodality approach [M]. New York:  Thieme Medical Publishers, 2011:  109.
[14] GRISWOLD M A,JAKOB P M, HEIDEMANN R M, et al. Generalized autocalibrating partially parallel acquisitions (GRAPPA) [J]. Magnetic Resonance Imaging,2002,47(6) : 1202-1210.
[15] BERNSTEIN M A,ZHOU X H J, POLZIN J A, et al. Concomitant gradient terms in phase contrast MR: analysis and correction [J]. Magnetic Resonance in Medicine,1998,39(2): 300-308.