Please wait a minute...
J4  2010, Vol. 44 Issue (2): 294-299    DOI: 10.3785/j.issn.1008-973X.2010.02.015
计算机技术﹑电信技术     
基于双波长激光扫描的皮下静脉显示装置
张帆, 刘向东
(浙江大学 现代光学仪器国家重点实验室,浙江 杭州 310027)
Subcutaneous vein display equipment based on dual-wavelength laser scanning
ZHANG Fan, LIU Xiang-dong
(State Key Laboratory of Modern Optical Instrumentation, Zhejiang University, Hangzhou 310027, China)
 全文: PDF  HTML
摘要:

阐述了将激光扫描显示技术应用于医疗辅助显示领域的优势.针对临床诊疗中穿刺操作因病患静脉血管不清晰而带来的困难,设计并搭建一种基于双波长激光扫描的皮下静脉血管显示装置,介绍了该装置的基本原理、系统构成、关键器件及工作方式.分别从静态和动态两方面对静脉血管与周围组织对红外光的吸收差异进行测量,结果表明,静脉与其他组织对红外光的吸收率差一般不超过10%.设计散射光收集装置,解决了到达探测面光能量过低以及扫描光点运动带来的干扰问题.所搭建的装置能够将红外光吸收较明显的静脉血管用可见激光显示在体表原有位置,达到了预期的实验效果.

Abstract:

Advantages of laser scanning display in the field of medical treatment aided display were introduced. Subcutaneous vein display equipment based on dual-wavelength laser scanning was designed and constructed due to the inconvenience of centesis in the clinic treat. The basic principle, structure, key components and working modes of the equipment were given. The infrared absorbance difference between vein and other adjacent tissue was measured statically and dynamically. Results showed that the difference of infrared absorbency was below 10%. A unit for collecting scattered light was built. Then the problems of comparatively low light energy arriving at the detector and the interference due to the movement of scanning light spot were overcomed. The subcutaneous vein having obvious infrared absorption can be displayed on the correct position by the equipment. The expectant result was achieved.

出版日期: 2010-03-09
:  TN 141.8  
通讯作者: 刘向东,男,教授.     E-mail: xdliu@zju.edu.cn
作者简介: 张帆(1984—),男,陕西宝鸡人,硕士生,从事光电投影显示的研究.
服务  
把本文推荐给朋友
加入引用管理器
E-mail Alert
RSS
作者相关文章  

引用本文:

张帆, 刘向东. 基于双波长激光扫描的皮下静脉显示装置[J]. J4, 2010, 44(2): 294-299.

ZHANG Fan, LIU Xiang-Dong. Subcutaneous vein display equipment based on dual-wavelength laser scanning. J4, 2010, 44(2): 294-299.

链接本文:

http://www.zjujournals.com/eng/CN/10.3785/j.issn.1008-973X.2010.02.015        http://www.zjujournals.com/eng/CN/Y2010/V44/I2/294

[1]  公丕欣,张静,宋明娇,等.疑难浅静脉穿刺技术与方法研究进展[J].现代护理,2004,10(11): 10541055.
GONG Pei-xin, ZHANG Jing, SONG Ming-jiao, et al. Research progress on centesis of difficult shallow subcutaneous veins[J]. Modern Nursing, 2004, 10(11): 10541055.
[2] ZEMAN H D, LOVHOIDEN G, VRANCKEN C. Prototype vein contrast enhancer[C]//Advanced Biomedical and Clinical Diagnostic Systems II, Proceedings of SPIE. Bellingham, WA: SPIE, 2004: 3949.
[3] ZEMAN H D, LOVHOIDEN G, VRANCKEN C, et al. Prototype vein contrast enhancer[J]. Optical Engineering, 2005, 44(8): 19.
[4] ZEMAN H D, LOVHOIDEN G. Enhancing the contrast of subcutaneous veins[C]// Biomedical Diagnostic, Guidance, and Surgical-Assist Systems,Proceedings of SPIE. California: SPIE, 1999: 219230.
[5] VILLASENOR C, SANCHEZ F J, GARAY M E. Contrast enhancement of mid and far infrared images of subcutaneous veins[J]. Infrared Physics and Technology, 2008, 51: 221228.
[6] SPRAGUE R, MOTAGUE T, BROWN D. Bi-axial magnetic drive for scanned beam display mirrors[C]//MOEMS Display and Imaging Systems III, Proceedings of SPIE. Bellingham, WA: SPIE, 2000: 113.
[7] 苗永平, 刘永智.二维激光扫描规律研究[J].应用光学,2005,26(2): 2730.
MIAO Yong-ping, LIU Yong-zhi. Investigation on laws of 2-D laser scanning[J]. Applied Optics, 2005, 26 (2): 2730.
[8] ZEMAN H D, LOVHOIDEN G, DESHMUKH H. Optimization of subcutaneous vein contrast enhancement[C]//Biomedical Diagnostic, Guidance, and Surgical-Assist Systems II, Proceedings of SPIE. California: SPIE, 2000: 5057.
[9] 郁道银, 谈恒英. 工程光学[M]. 北京:机械工业出版社,1999: 67.

No related articles found!