Please wait a minute...
浙江大学学报(工学版)
机械与电气工程     
磁共振电特性成像人体腹腔癌症仿真研究
刘阳, 于孝辉, 郭建炎
1. 厦门理工学院 电气工程与自动化学院,福建 厦门 361024
2. 休斯顿卫理公会派医院研究中心 系统医药与生物工程系,德克萨斯 休斯顿 77030
3. 厦门理工学院 光电与通信工程学院,福建 厦门 361024
Simulation study on magnetic resonance electrical property tomography of human abdominal cancer
LIU Yang, YU Xiao hui, GUO Jian yan
1. School of Electrical Engineering and Automation, Xiamen University of Technology, Xiamen 361024, China;
2. Department of Systems Medicine and Bioengineering, Houston Methodist Hospital Research Institute, Houston 77030, USA;
3. School of Opto-Electronic and Communication Engineering, Xiamen University of Technology, Xiamen 361024, China
 全文: PDF(2144 KB)   HTML
摘要:

将磁共振电特性成像(MREPT)方法用于人体腹腔部位,考察该成像方式用于人体腹腔肝癌及胰腺癌诊断的可行性.采用鸟笼线圈同时作为射频发射和接收,基于射频发射场测量值重构人体腹腔内组织的电导率和相对介电率.包含10种组织的人体腹腔模型上的仿真结果表明:在信噪比SNR=30的情况下,MREPT能够检测出位于组织内部半径为3 cm的肝癌肿瘤和半径为1 cm的胰腺癌肿瘤;由于逆问题重构公式中均质电特性的假设及拉普拉斯卷积的共同作用,可能无法检测出位于组织边界处的小体积癌症肿瘤.拉普拉斯卷积核的大小对重构结果的影响很大,在无噪声情况下,3×3×3的卷积核的重构效果最好,但当SNR=30~120时,5×5×5的卷积核是最佳选择.

Abstract:

The magnetic resonance electrical property tomography (MREPT) method was applied to human abdomen. The birdcage coil was used for radio frequency (RF) transmission and reception. The conductivity and relative permittivity of abdominal tissues were reconstructed based on the RF field measurement. The simulation results conducted on human abdominal model including 10 types of tissues show that the MREPT can detect the 3cm-radius liver cancer tumor and 1cm-radius pancreas tumor located in the central area of tissues. MREPT may fail to detect the small tumors located near the boundary of tissues due to the homogeneous electrical property assumption made in the reconstruction algorithm together with the adopted Laplacian revolution. The size of Laplacian revolution kernel has significant influence on the reconstruction result. When there is no noise, the 3×3×3 Laplacian revolution kernel can achieve the highest accuracy. The best performance can be achieved by 5×5×5 Laplacian revolution kernel with a SNR in the range of 30 to 120.

出版日期: 2016-10-28
:  TM 15  
基金资助:

福建省自然科学基金资助项目(2013J05102,2016J01323); 福建省教育厅科研资助项目(JA15383).

通讯作者: 郭建炎, 男, 副教授. ORCID: 0000-0002-3508-0755.     E-mail: guonetmail@163.com
作者简介: 刘阳(1980—), 女, 讲师, 博士,从事生物电阻抗成像的研究. ORCID: 0000-0001-8631-0154. E-mail: liuyang32201220@gmail.com
服务  
把本文推荐给朋友
加入引用管理器
E-mail Alert
RSS
作者相关文章  

引用本文:

刘阳, 于孝辉, 郭建炎. 磁共振电特性成像人体腹腔癌症仿真研究[J]. 浙江大学学报(工学版), 10.3785/j.issn.1008-973X.2016.10.016.

LIU Yang, YU Xiao hui, GUO Jian yan. Simulation study on magnetic resonance electrical property tomography of human abdominal cancer. JOURNAL OF ZHEJIANG UNIVERSITY (ENGINEERING SCIENCE), 10.3785/j.issn.1008-973X.2016.10.016.

链接本文:

http://www.zjujournals.com/eng/CN/10.3785/j.issn.1008-973X.2016.10.016        http://www.zjujournals.com/eng/CN/Y2016/V50/I10/1959

[1] KESHTKAR A, SALEHNIA Z, KESHTKAR A, et al. Bladder cancer detection using electrical impedance technique (Tabriz Mark1) [J]. Pathology ResearchInternational, 2012, 2012(2012): 470101.
[2] WALKER D C, SMALLWOOD R H, KESHTKAR A, et al. Modeling the electrical properties of bladder tissuequantifying impedance changes due to inflammation and oedema [J]. Physiological Measurement, 2005,26(3): 251-268.
[3] POPLACK S P, TOSTESON T D, WELLS W A, et al. Electromagnetic breast imaging: results of a pilot study in women with abnormal mammograms [J]. Radiology, 2007, 243(2): 350-359.
[4] HALTER R J, SCHNED A, HEANEY J, et al. Electrical properties of prostatic tissues: I. single frequency admittivity properties [J]. The Journal of Urology, 2009, 182(4): 1600-1607.
[5] ROURKE A P, LAZEBNIK M, BERTRAM J M, et al. Dielectric properties of human normal, malignant and cirrhotic liver tissue: in vivo and ex vivo measurements from 0.5 to 20 GHz using a precision openended coaxial probe [J]. Physics in Medicine and Biology, 2007, 52(15): 4707-4719.
[6] KESHTKAR U, VOIGT T, FINDEKLEE C, et al. Determination of electric conductivity and local SAR via B1 mapping [J]. IEEE Transaction on Medical Imaging, 2009, 28(9): 1365-1374.
[7] BULUMULLA S B, LEE S K, YEO D T B. Conductivity and permittivity imaging at 3.0T [J]. Concepts in Magnetic Resonance Part B: Magnetic Resonance Engineering, 2012, 41B(1): 13-21.
[8] WAN Y, HALTERR, BORSIC A, et al. Sensitivity study of an ultrasound coupled transrectal electricalimpedance tomography system for prostate imaging [J]. Physiological Measurement, 2010, 31(8): 17-29.
[9] JEON K, KIM H J, LEE C, et al. Integration of the denoising, inpainting and local harmonic Bz algorithm for MREIT imaging of intact animals [J]. Physics in Medicine and Biology, 2010, 55(24): 7541-7548.
[10] OH T I, JEONG W C, MCEWAN A, et al. Feasibility of magnetic resonance electrical impedance tomography (MREIT) conductivity imaging to evaluate brain abscess lesion: in vivo canine model [J]. Journal of Magnetic Resonance Imaging, 2013, 38(1): 189-197.
[11] VOIGT T, KESHTKAR U, DOESSEL O. Quantitative conductivity and permittivity imaging of the human brain using electric properties tomography [J]. Magnetic Resonance in Medicine, 2011, 66(2): 456-466.
[12] ZHANG Xiaotong, VAN DE MOORTELE P, SCHMITTER S, et al. Complex B1 mapping andelectrical properties imaging of the human brain using a 16channel transceiver coil at 7T [J]. Magnetic Resonance in Medicine, 2013, 69(5): 1285-1296.
[13] HAACKE E M, PETROPOULOS L S, NILGES E W, et al. Extraction of conductivity and permittivity using magnetic resonance imaging [J]. Physics in Medicine and Biology, 1991, 36(6): 723-734.
[14] WEN Han. Noninvasive quantitative mapping of conductivity and dielectric distributions using the RF wave propagation effects in high field MRI [C]∥Medical Imaging 2003: Physics of Medical Imaging. San Diego: [s. n.], 2003, 5030: 471-477.
[15] VAN LIER A L H M W, BRUNNER D O, PRUESSMANN K P, et al. B1+ phase mapping at 7T and its application for in vivo electrical conductivity mapping [J]. Magnetic Resonance in Medicine, 2012, 67(2): 552-561.
[16] VAN LEER A L H M W, RAAIJMAKERS A, VOIGT T, et al. Electrical properties tomography in the human brain at 1.5, 3 and 7T: a comparison study [J]. Magnetic Resonance in Medicine, 2014, 71(2): 354-363.
[17] LIU Jiaen, ZHANG Xiaotong, MOORTELE P, et al. Determining electrical properties based on B1 fields measured in an MR scanner using a multi-channel transmit/receive coil: a general approach [J]. Physics in Medicine and Biology, 2013, 58(13): 4395-4408.
[18] BALIDEMAJ E, VAN LEER A L H M W, CREZEE H, et al. Feasibility of electric property tomography of pelvic tumors at 3T [J]. Magnetic Resonance in Medicine, 2015, 73(4): 15051513.
[19] SHIN J, KIM M J, LEE J, et al. Initial study on invivo conductivity mapping of breast cancer using MRI [J]. Journal of Magnetic Resonance Imaging, 2014,42(2): 371378.
[20] WANG Jinghua, QIU Maolin, YANG Qing, et al. Measurement and correction of transmitter and receiver induced nonuniformities in vivo [J]. Magnetic Resonance in Medicine, 2005, 53(2): 408-417.
[21] YARNYKH V L. Method for rapid threedimensional mapping of the transmitted radiofrequency field [J]. Magnetic Resonance in Medicine, 2007, 57(1): 192-200.
[22] HASGALL P A, NEUFELD E, GOSSELIN M C, et al. It is database for thermal and electromagneticparameters of biological tissues [J]. Journal of Cell Biology, 2011, 93(1): 170.
[23] RAOOF M, CISNEROS B T, CORR S J, et al. Tumor selective hyperthermia induced by shortwave capacitivelycoupled RF electricfield [J]. PloS ONE, 2013, 8(7): 68506.
[24] SAVITZKY A, GOLAY M J. Smoothing and differentiation of data by simplified least square procedures [J]. Analytical Chemistry, 1964, 36(8): 1627-1639.

[1] 刘阳, 戴松世. 注入及感应电流磁共振电阻抗成像对比[J]. J4, 2011, 45(10): 1848-1854.
[2] 刘阳,吴占雄,朱善安,贺斌. 各向异性头部组织感应电流磁共振电阻抗成像[J]. J4, 2011, 45(1): 168-172.
[3] 王启涵, 姚缨英. 电磁感应加热中控制电压的计算[J]. J4, 2010, 44(8): 1553-1557.
[4] 孟豹, 黄平捷, 叶波, 等. 脉冲涡流解析模型研究范[J]. J4, 2009, 43(09): 1621-1624.