浙江大学学报(工学版)  2017, Vol. 51 Issue (10): 2077-2083    DOI: 10.3785/j.issn.1008-973X.2017.10.024
电信技术     
基于高速摄像机的动态血压非接触获取
许文媛, 孟濬, 赵夕朦
浙江大学 电气工程学院, 浙江 杭州 310027
Non-contact monitoring of ambulatory blood pressure based on high speed camera
XU Wen-yuan, MENG Jun, ZHAO Xi-meng
College of Electrical Engineering, Zhejiang University, Hangzhou 310027, China
 全文: PDF(2155 KB)   HTML
摘要:

为了实现基于高速摄像机的非接触、非介入式的动态血压获取,对血压与脉搏信号相位差、心率之间的函数关系进行理论推导;基于成像式光电容积描记技术,实现了通过高速摄像机非接触获取脉搏信号的相位差和心率;基于血压与脉搏信号相位差、心率之间的函数关系以及基于高速摄像机所获得的脉搏信号相位差和心率,通过曲线拟合得到血压与脉搏信号相位差、心率之间的函数关系的参数值,实现了基于高速摄像头的动态血压的非接触获取.为了验证基于高速摄像头获得的血压的准确性,以欧姆龙智能电子血压仪获得的血压为参照进行误差分析,所得的收缩压的误差为-4.03%~8.44%,舒张压的误差为-6.58%~8.90%.

Abstract:

A non-contact and non-invasive acquisition of ambulatory blood pressure was achieved based on a high speed camera. The functional relationship among blood pressure, phase difference of pulse wave and heart rate was obtained through theoretical deduction. Phase difference of pulse wave and heart rate were acquired through the high speed camera based on imaging photoplethysmographic. Parameters of the functional relationship were determined through curve fitting based on the functional relationship and the acquired phase difference and heart rate. Ambulatory blood pressure was obtained through the high speed camera. Error analysis was processed with reference to an Omron intelligent electronic sphygmomanometer in order to verify the accuracy of blood pressure acquired through the high speed camera. The obtained error of systolic blood pressure varied from -4.03 to 8.44 percent. The obtained error of diastolic blood pressure varied from -6.58 to 8.90 percent.

收稿日期: 2016-09-01 出版日期: 2017-09-27
CLC:  TN911  
基金资助:

浙江省公益性技术应用研究计划资助项目(2017C31079);浙江省科技厅科技计划公益技术研究基金资助项目(2011C23097);浙江省教育厅资助项目(Y20107866).

通讯作者: 孟濬,男,副教授.ORCID:0000-0002-7633-3624.     E-mail: junmeng@zju.edu.cn
作者简介: 许文媛(1992-),女,硕士生,从事生物特征提取和智能感知的研究.ORCID:0000-0002-0813-7138.E-mail:21410183@zju.edu.cn
服务  
把本文推荐给朋友
加入引用管理器
E-mail Alert
作者相关文章  

引用本文:

许文媛, 孟濬, 赵夕朦. 基于高速摄像机的动态血压非接触获取[J]. 浙江大学学报(工学版), 2017, 51(10): 2077-2083.

XU Wen-yuan, MENG Jun, ZHAO Xi-meng. Non-contact monitoring of ambulatory blood pressure based on high speed camera. JOURNAL OF ZHEJIANG UNIVERSITY (ENGINEERING SCIENCE), 2017, 51(10): 2077-2083.

链接本文:

http://www.zjujournals.com/eng/CN/10.3785/j.issn.1008-973X.2017.10.024        http://www.zjujournals.com/eng/CN/Y2017/V51/I10/2077

[1] 王继光.24 h动态血压监测:适用人群及临床意义[J].中华高血压杂志,2014, 22(7):614-615. WANG Ji-guang. Twenty-four-hour ambulatory blood pressure monitoring:applicable crowd and its clinical significance[J]. Chinese Journal of Hypertension, 2014, 22(7):614-615.
[2] O'BRIEN E, PARATI G, STERGIOU G, et al. European society of hypertension position paper on ambulatory blood pressure monitoring[J]. Journal of Hypertension, 2013, 31(9):1731-1768.
[3] O'BRIEN E. Twenty-four-hour ambulatory blood pressure measurement in clinical practice and research:a critical review of a technique in need of implementation[J]. Journal of Internal Medicine, 2011, 269(5):478-495.
[4] 非接触、非介入式改变医疗保健[DB/OL].(2013-07-25)[2016-08-28].http://news.cecb2b.com/info/20130725/920756.shtml.
[5] Non-contact and non-invasive changes medical insurance.[DB/OL]. (2013-07-25)[2016-08-28]. http://news.cecb2b.com/info/20130725/920756.shtml.
[6] 张政波,吴太虎.无创血压测量技术与进展[J].中国医疗器械杂志,2003,27(3):196-199. ZHANG Zheng-bo, WU Tai-hu. Noninvasive blood pressure measurement technology and its progress[J]. Chinese Journal of Medical Instrumentation, 2003,27(3):196-199.
[7] 郑理华,窦建洪,何兴华,等.无创血压测量技术的改进与进展[J].中国医学装备,2013, 10(3):49-52. ZHENG Li-hua, DOU Jian-hong, HE Xing-hua, et al. Improvement and progress of noninvasive blood pressure measurement technology[J]. China Medical Equipment, 2013, 10(3):49-52.
[8] MCCOMBIE D B, SHALTISP A, REISNER A T, et al. Adaptive hydrostatic blood pressure calibration:Development of a wearable, autonomous pulse wave velocity blood pressure monitor[C]//200729th Annual International Conference of the IEEE Engineering in Medicine and Biology Society. Lyon:IEEE, 2007:370-373.
[9] 李申龙,李毅彬,李洪阳,等.基于脉搏波相位差的无创连续血压测量方法[J].传感器与微系统,2016, 35(1):62-64. LI Shen-long, LI Yi-bin, LI Hong-yang, et al. Non-invasive continuous blood pressure measurement based on phase difference of pulse wave[J]. Transducer and Microsystem Technologies, 2016, 35(1):62-64.
[10] 向海燕,俞梦孙.用脉搏波传导时间实现血压连续测量[J].医疗卫生装备,2006,27(2):19-21. XIANG Hai-yan, YU Meng-song. Measurement of continuous blood pressure via pulse wave translation time[J]. Medical Equipment, 2006, 27(2):19-21.
[11] ZAKARIA N A, SHARIFMUDDIN N B, RIDZWAN W M F W M, et al. Pulse wave transit time and its relationship with systolic blood pressure[C]//6th World Congress of Biomechanics (WCB 2010). Singapore:Springer, 2010:1354-1357.
[12] SUN Y, PAPIN C, AZORIN-PERIS V, et al. Use of ambient light in remote photoplethysmographic systems:comparison between a high-performance camera and a low-cost webcam[J]. Journal of Biomedical Optics, 2012, 17(3):0370051-03700510.
[13] SUN Y, HU S, AZORIN-PERIS V V, et al. Noncontact imaging photoplethysmography to effectively access pulse rate variability[J]. Journal of Biomedical Optics, 2013, 18(6):061205.
[14] 李章俊,王成,朱浩,等.基于光电容积脉搏波描记法的无创连续血压测量[J].中国生物医学工程学报,2012,31(4):607-614. LI Zhang-jun, WANG Cheng, ZHU Hao, et al. Non-invasive continuous blood pressure measurement based on photoplethysmographic[J]. Chinese Journal of Biomedical Engineering, 2012, 31(4):607-614.
[15] KAMSHILIN A A, MIRIDONOV S, TEPLOV V, et al. Photoplethysmographic imaging of high spatial resolution[J]. Biomedical Optics Express, 2011, 2(4):996-1006.
[16] TENG X F, ZHANG Y T. The effect of applied sensor contact force on pulse transit time[J]. Physiological Measurement, 2006, 27(8):675-684.
[17] TENG X F, ZHANG Y T. The effect of contacting force on photoplethysmographic signals[J]. Physiological Measurement, 2004, 25(5):1323-1335.
[18] JONATHAN E, LEAHY M J. Cellular phone-based photoplethysmographic imaging[J]. Journal of Biophotonics, 2011, 4(5):293-296.
[19] JONATHAN E, LEAHY M J. Investigating a smartphone imaging unit for photoplethysmography[J]. Physiological Measurement, 2010, 31(11):N79.
[20] POH M Z, MCDUFF D J, PICARD R W. Advancements in noncontact, multiparameter physiological measurements using a webcam[J]. IEEE Transactions on Biomedical Engineering, 2011, 58(1):7-11.
[21] HUMPHREYS K, WARD T, MARKHAM C. Noncontact simultaneous dual wavelength photoplethysmography:a further step toward noncontact pulse oximetry[J]. Review of Scientific Instruments, 2007, 78(4):044304.
[22] 肖晓.血压为何会升高[J].药物与人,2004(1):14. XIAO Xiao. Why blood pressure rises[J]. Medicine and People, 2004(1):14.

[1] 余建波, 李传锋, 吕靖香. 轴承故障信号的平均组合差值形态滤波分析[J]. 浙江大学学报(工学版), 2018, 52(10): 1845-1853.
[2] 郝静, 杜太行, 江春冬, 陈通海. 调参级联随机共振系统加强策略[J]. 浙江大学学报(工学版), 2017, 51(10): 2084-2092.
[3] 彭岑, 赵民建, 郑雅敏. TCM-8PSK短包高速跳频接收机的快速载波同步[J]. J4, 2010, 44(4): 625-631.
[4] 顾宇杰, 史治国, 李钰, 等. 基于二次采样的贝叶斯波束成形[J]. J4, 2009, 43(5): 812-816.